9-5 OP-07 ANALOG OUTPUT
The OP-07 analog output option is for sending the weight data to an analog input unit. The output is a 4 to 20 mA
current output proportional to the display reading.
The output data is updated in synchronization with the display update.
Setting
ANF- Name Default
Parameter Description
1 Same as display
2 Gross
01 Output data 1
3 Net
02
Weight
value at 4
mA output
0
03
Weight
value at 20
mA output
16000
-999999
to
9999999
The position of decimal point is set in CALF-02.
Interface Specification
Current output 4 to 20 mA
Applicable load resistance
0 to 520 Ω
Resolution Approximately 1 / 3000
Non-linearity +/- 0.1% F.S or less
Zero point: +/- 0.02 %FS/ °C or less
Temperature coefficient
Sensitivity: +/- 0.02 %FS/ °C or less
SETTING METHOD
Set the weight value at a 4 mA output point in ANF-02 and that at a 20 mA output
point in ANF-03, ignoring the unit of weight.
<Setting Example>
Outputting 4 mA at 0.0 kg and 20
mA at 1,000.0 kg. (Full line in the
figure at the left )
ANF-02 = 0.0 kg
ANF-03 = 1000.0 kg
Outputting 20 mA at 0.0 kg and 4 mA at 1,000.0 kg. (Dotted line in the
figure at the left)
ANF-02 = 1000.0 kg
ANF-03 = 0.0 kg