CALIBRATION RELATED FUNCTIONS
Setting
CALF- Name
Default Parameter Description
0 None
1 g (International version)
2 kg ( International version and USA version)
3 t (International version)
01 Weighing Unit 2
4 lb. ( USA version)
0 None 12345
1
10
1
1234.5
2
10
2
123.45
3
10
3
12.345
02
Decimal Point
Position
0
4
10
4
1.2345
03
Minimum
Division
1
1, 2, 5, 10,
20, or 50
Minimum division (increment) for the weight value.
Input 1,2, 5, 10, 20, or 50(decimal point ignored).
04 Capacity 16000 0 to 800000
Capacity of the weighing. Weighing is allowed up to this
setting + 9D(9 divisions). A weight value beyond this limit
is an overflow and not displayed.
05 Zero Range 2 0 to 30
A range of accepting "zero" from the [ZERO] key or
Control I/O. Represented in terms of percent(&) with
respect to the weighing capacity, centering around the zero
calibrated point in calibration. If this is set to 2, "zero" can
be accepted within a range of the zero calibrated point
+/- 2 %.
06
Zero Tracking
Time
0.0 0.0 to 5.0
Zero tracking is performed in combination with CALF-07
Zero Tracking Width. It is not performed when set to 0.0.
Its unit is seconds
Zero Tracking
Width
0 0 to 9
Zero tracking is performed in combination with CALF-06 Zero
Tracking Time. It is not performed when set to 0. Its least input
increment is 1/2D. (The width for setting of 1 is equivalent to 1/2 of
the minimum graduation)
07