69
Type of Response Description
Acknowledgment
Communication procedure A: Repeats the command as it is.
Communication procedure B: Returns the data.
Communication procedure C: Returns the command as it is
and waits to receive the data.
Negative acknowledgment (Beyond condition): AD-4401 in
non-receiving mode
IE
Negative acknowledgment (Beyond range):
Data format correct, but value beyond allowable range
VE
Negative acknowledgment (Illegal command/data): Command or
data format illegal
?E
Any response is made within 200 ms after receiving the command.
Command Name Function
RW Request-Weight
Reads out the weight; The content of the response data is the same as that set with
RSF-01. Same operation as when the manual print key is pressed.
Data is sent in the “A&D standard format”.
RB Request-Batch-status
Reads out the weighing sequence status; The response data is sent in the
“Output-per-Sampling Format”. The weight output value is the net.
RF Request-Final
Reads out the weighing (Batching) result; Outputs net weight when weighting completes.
Data is sent in the “A&D standard format”.
RT Request-Total
Reads out the accumulated weight and times;
Data is sent in the “Accumulation data format”.
Set Setpoints
Sets the setpoints; Writes 42-character for 7 setpoints altogether. Blanks are filled with 0.
Each value is shown in 6 digit without decimal point. Data is sent in the following format.
SS
RS Request-Setpoint Reads out the Setpoints.; Same format as SS
RE Read-EEPROM
Reads from the EEPROM; Reads the calibration or function data from the EEPROM. The
data equivalent to 206 characters is returned in the ASCII code except the terminators.
If this data is written to another AD-4401 with a “WE” command, calibration and function
settings can be copied.
WE Write-EEPROM Writes to the EEPROM: Same format as the RE command