59
<ACCUMULATION DATA FORMAT>
< Example >
Header-
1
8 Digits Including Data, Polarity, and Decimal Point Unit Terminator
Accumulated
weight
T W , + 0 1 2 3 4 5 6 . 7 8 k g CR LF
Header
is "TW"
Weight +
Overflow
T W , + SP SP SP SP SP SP SP . SP SP k g CR LF
"SP"
value
Weight -
Overflow
T W , - SP SP SP SP SP SP SP . SP SP k g CR LF
"SP"
value,
"-"
polarity
Accumulated
count
T N , + 0 1 2 3 4 5 6 7 8 9 SP SP CR LF
Header
is "TN"
Counts +
Overflow
T N , + SP SP SP SP . SP SP . SP SP SP SP CR LF
"SP"
value
[Stream]
The data is transmitted in synchronization with the update of the display. When the display rewrite exceeds the baud
rate, data transmission is suspended until the next display update.
The transmitted data is the same as what is being displayed. Therefore, non-displayed data is never transmitted.
[Auto Print]
The data is transmitted each time weighing has been completed. The weighing completion timing differs depending on
the weighing mode. See CHAPTER 7 WEIGHING SEQUENCE.
[Manual Print]
The data is transmitted when you press the key set for the “print command for manual print” or a Control I/O input is
made.