7-6 Error Codes
When the instrument detects an erroneous measurement condition, the following error codes are
displayed.
Error code Symptom Remedy
Blood pressure related errors
E00
Zero point error in the pressure
sensor circuit
Exhaust air from the cuff and turn on the power
again.
E11 Can not pressurize.
E12 Pressurizing speed is too slow.
Check the cuff and hose for correct connection or
check them for being folded.
E13 Pressurizing speed is too fast.
Confirm that the patient setting is correct. (Adult,
Neonate)
E21
Measurement time is too long.
Constant exhaustion speed is
too slow.
E22 Exhaustion speed is too fast.
E23 Excessive pressure is detected.
Check the cuff and hose for correct connection or
check them for being folded.
E42 Insufficient pressurization
E43 No pulse is detected.
E44 Physical movement is detected.
E45
The diastolic blood pressure
cannot be determined.
E46
The mean blood pressure
cannot be determined.
E48
The systolic blood pressure
cannot be determined
E61
The pulse rate cannot be
determined.
E63
The blood pressure value is
inappropriate.
Check the cuff for correct placement or check the
patient for physical movement or for irregular pulse.
SpO
2
related errors
E30 SpO
2
self test error Turn on the power again.
E31 SpO
2
sensor error Confirm that the specified sensor is connected.
ECG related errors
E35 ECG self test error
E36 Defective ECG circuit
Turn on the power again.
Other error
E80 System error Turn on the power again.
Note
• Cuffs and sensors are consumables. If a measurement error occurs frequently, they must be replaced.
Refer to "8 Accessories/Options" for detail information.
• When the optional expansion box is connected, error codes other than listed in the table may appear.
Refer to the expansion box instruction manual for details.