42
7. Maintenance
7-1 Cleaning
Caution
• Before cleaning, turn off the power and remove the AC adapter.
• Do not pour water on the instrument or use water for cleaning. This instrument is not
waterproof.
• Do not sterilize the instrument by gaseous sterilization or using an autoclave.
• Do not use solvents such as thinner and benzine for cleaning. That may deform or
discolor the instrument.
Clean the instrument as described below:
Clean the instrument exterior with a dry soft cloth.
If the instrument is stained with blood, chemical substance, or dirt, wipe the instrument off using a cloth
dampened with a mild detergent.
For cleaning a patient's body site to apply a sensor, follow the procedure of each hospital.
To disinfect the cuff, wipe the surface which comes into contact with the skin with alcohol such as 70%
isopropyl. Do not wash the cuff.
Discard the disposables after use. To disinfect the optional SpO
2
sensor or the extension cable, use
alcohol such as 70% isopropyl or bleach diluted by 10%.
Wipe the ECG extension cable and ECG patient cable with alcohol such as 70% isopropyl.