41
No. Setting item Setting range Description
S00 Station address 0-98 (0) To connect an option.
S10 Communication mode 1-10 (1)
To connect an option.
To be selected, depending on the option.
S11 Baud rate
1200, 2400, 4800,
9600
, 19200, 38400
To connect an option.
To be selected, depending on the option
S12 Data bit 7, 8
To connect an option.
To be selected, depending on the option
S13 Stream mode OFF, ON
To connect an option.
To be selected, depending on the option
S21 Previous data hold OFF, ON
Whether to retain the previous data or not
when the POWER switch is turned off.
"ON" to retain.
S22 Averaging pulse 1, 4, 8
Sets the number of ECG pulse rate data
for moving average.
S23
Just in time
measurement
OFF, ON
Whether to use an exact-time measurement
for blood pressure or not.
"ON" to use.
S24 Sensor removal alarm OFF, ON
Whether to sound an alarm when the
sensor removal is detected.
"OFF" not to sound.
S25 Heart rate -pulse
AUTO
, ECG, SpO
2
,
NIBP
Specifies whether to display, heart rate or
pulse rate and pulse sound.
"AUTO" to sound pulse in synchronization
with the pulse detected according to the
order of precedence.
Underlined items: Default settings