- 48 - AD-4430B
7.6.5. The List Of The Linearization Function
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting
MEMO
Lnr01
1 to 4
Number of
input points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the
set value is between 1 and 2.
0
Lnr02
-7.0000
to 7.0000
Linear-zero Voltage for linear-zero input. Unit : mV/V. 0.0000
Lnr03
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on CAL02.
0
Lnr04
0.0000
to 9.9999
Linear 1
Span
The span voltage between linear-zero and linear
1 input. Unit : mV/V.
0.0000
Lnr05
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on CAL02.
0
Lnr06
0.0000
to 9.9999
Linear 2
Span
The span voltage between linear-zero and linear
2 input. Unit : mV/V.
0.0000
Lnr07
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on CAL02.
0
Lnr08
0.0000
to 9.9999
Linear 3
Span
The span voltage between linear-zero and linear
3 input. Unit : mV/V.
0.0000
Lnr09
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on CAL02.
0
Lnr10
0.0000 to
9.9999
Linear 4
Span
The span voltage between linear-zero and linear
4 input. Unit : mV/V.
0.0000