AD-4430B - 47 -
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting
MEMO
CAL12
0 to 1
Output when
overflow and
unstable
Standard serial output when the weight value
overflows and is unstable.
0 : Disables output.
1 : Enables output.
1
CAL13
1 to 3
Exceeding
negative
gross weight
To judge when the negative gross weight is
exceeded.
1 : Gross weight < -99999
2 : Gross weight < Negative weighing
capacity
3 : Gross weight < -19d
1
CAL14
1 to 2
Exceeding
negative net
weight
To judge when the negative net weight is
exceeded.
1 : Net weight < -99999
2 : Net weight < Negative weighing capacity
1
CAL15
0 to 1
Clear the zero
value
Select whether or not to clear the zero value.
0 : Disables.
1 : Enables.
1
CAL16
0 to 1
Zero setting
when power
is turned on
Select whether or not to perform zero setting
when power is turned on.
0 : Disables.
1 : Enables.
0
CAL17
-7.0000
to 7.0000
Input voltage
at zero
Input voltage from a load cell at zero. Unit :
mV/V.
This value is determined in zero calibration
during the calibration with an actual load.
0.0000
CAL18
0.0001
to 9.9999
Input voltage
at span
Input voltage from a load cell at span. Unit :
mV/V.
This value and the value of CAL19 are
determined in span calibration during the
calibration with an actual load.
3.2000
CAL19
1 to 99999
Calibration
weight value
correspondin
g to Input
voltage at
span
The calibration weight value corresponding to
the Input voltage at span of CAL18. When
performing digital span, CAL17, CAL18 and
CAL19 are required. The decimal point
position is the same as the setting of CAL02.
32000
CAL26
9.7500
to 9.8500
Gravity
acceleration
of place of
calibration
Gravity acceleration of the place of calibration.
Unit : in m/s
2
.
9.8000
CAL27
9.7500
to 9.8500
Gravity
acceleration
of place of
use
Gravity acceleration of the place where the
scale is being used. Unit : in m/s
2
.
9.8000
CAL28
0 to 1
Disable
averaging
hold
0 : Disables.
1 : Enables.
0