A&D AD-4430B Specifikace Strana 24

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 56
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 23
- 22 - AD-4430B
Illustration 8 Gravity Acceleration Graph
5.3.4. Outline Of The Linearization Function
Even if zero and span points have been calibrated, weighing errors may occur between the
zero point and weighing capacity. This is a corrective function designed to non-linearly correct
weighing errors.
It is possible to input up to four points in addition to the zero point.
The zero point and each input point will be corrected to put them in a straight line.
Areas between input points that could not be corrected completely with straight line
correction or with secondary correction will be corrected using a curved line derived from
high-order equations.
When the actual load input for digital linearization is performed, the calibrated data will be
refreshed using zero point and final input point data. It is not necessary to calibrate again.
Illustration 9 Digital linearization
Zero
1s
t
4th3rd2nd
Mass value (Input point)
Weighing value
Before correction
Straight line correction
High-order equation correction
Zobrazit stránku 23