- 44 - AD-4430B
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting
MEMO
Fnc09
1 to 2
Mass to be
compared
with
near-zero
Item to be compared with near-zero.
1 : Gross weight
2 : Net weight
1
Fnc10
-99999 to
99999
Set value of
upper limit
Reference value for the upper limit.
Decimal point position is linked to CAL02.
10
Fnc11
-99999 to
99999
Set value of
lower limit
Reference value for the lower limit.
Decimal point position is linked to CAL02.
-10
Fnc12
1 to 2
Mass to be
compared
with upper
and lower
limit
Item to be compared with the upper and lower limit.
1 : Gross weight
2 : Net weight
1
Fnc13
1 to 2
Output logic
of upper
and lower
limit
Logic used when the result of the comparison with
the upper and lower limit is output
1 : Positive logic
2 : Negative logic
1
7.6.2. The List Of The Hold Function
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting
MEMO
HLd01
0.00 to 9.99
Averaging
time
Time to perform the averaging. Unit : second.
0.00 is not averaged.
0.00
HLd02
0.00 to 9.99
Start wait
time
Time to wait before commencing hold or averaging.
Unit : second.
0.00
HLd03
0 to 2
Condition of
automatic
start
Condition to start the hold or averaging automatically
.
0 : Do not use automatic start.
1 : Above the near-zero range, and stable.
2 : Above the near-zero range.
0
HLd04
0 to 1
Release
due to
falling
control input
Release when control input is falling.
0 : Do not release
1 : Release
1
HLd05
0.00 to 9.99
Time
release
Release after a set amount of time has passed.
Unit : second. 0.00 is not averaged.
0.00
HLd06
0 to 99999
Release
due to
range of
fluctuation
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is
the same as the setting of CAL02.
0
HLd07
0 to 1
Release in
near-zero
range
Release when the weighing value is in the
near-zero range.
0 : Do not release.
1 : Release.
0