AD-4430B - 25 -
Function No.
Parameter
and range
Function name Description
Default
value
CAL02
0 to 0.0000
Decimal point
position
Decimal point position of the weight value
0 0.000
0.0 0.0000
0.00
0
CAL03
1 to 50
Minimum division
Minimum division (d) of the weight value
1 10
2 20
5 50
1
CAL04
1 to 99999
Weighing capacity
Weighing capacity of the module.
Weighing is possible up to the value of this setting plus
8 d. If the value exceeds this, overflow will occur and
will not be displayed. The decimal point position is the
same as the setting of CAL02.
70000
CAL05
0 to 100
Zero adjustment
range
Range to enable zero adjustment by the (ZERO)
key Expressed as a percentage of the weighing
capacity with the calibration zero point as the center.
For example, if this is set to 2, the value in the range of
±2% of the weighing capacity with the calibration zero
point at the center will be to zero. When a power-ON
zero is performed, the initial zero point will be the
center.
2
CAL06
0.0 to 5.0
Zero tracking time
Performs zero tracking using this setting in
combination with the setting of the zero tracking band
CAL07. When 0.0, zero tracking will not be performed.
Unit : Second in 0.1 increments.
0.0
CAL07
0.0 to 9.9
Zero tracking band
Performs zero tracking using this setting in
combination with the setting of the zero tracking time
CAL06. When 0.0, zero tracking will not be performed.
Unit : d (minimum division) in 0.1 increments.
0.0
Displayed valu
Displayed value
0 d
4.5 d
When CAL06 = 1.0, CAL07 = 4.
1 second
Zero tracking follows the weight
value drifting a round the zero point
and adjusts to display as zero.
Zero tracking functions when the
weight value is drifting the range
shown in the graph.
1 second
2 second
5.0 d
4.5 d
4.0 d
3.5 d
3.0 d
2.5 d
2.0 d
1.5 d
1.0 d
0.5 d
0.0 d
When
CAL06 = 2.0,
CAL07 = 0.5
When
CAL06 = 1.0,
CAL07 = 4.5