A&D GH Series Manuál s instrukcemi Strana 40

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 71
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 39
A&D GR Series Instruction Manual
GB 40
KF format SiF tyPE 2
This is the Karl-Fischer moisture meter format, and is used when the peripheral equipment can only com-
municate using this format.
This format consists of fourteen characters (excluding the terminator).
This format has no header characters.
The polarity sign is first, if not zero or overloaded.
The weight data uses spaces in place of leading zeros.
This format outputs the unit "g" only for a stable value.
Data Unit
Terminator
Stable value
Unstable value
MT format SiF tyPE 3
This format has a two character header.
The polarity sign is used only for negative data.
The weight data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit.
Header
Data Unit
Terminator
Stable header
Unstable header
Overload header
NU (numerical) format SiF tyPE 4
This format has only numerical data.
This format consists of nine characters (excluding the terminator).
The polarity sign is first.
The weight data uses leading zeros.
Data
Terminator
Data number format dout d-no 1
This data number format is output just before data is transmitted to the RS-232C interface.
This format consists of six characters (excluding the terminator).
Data number
Terminator
Zobrazit stránku 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 70 71

Komentáře k této Příručce

Žádné komentáře