3.7. Correcting the Internal Weight Value
Note: The GR series can correct the internal weight value within 21.5 mg. The initial
internal weight value of the GR-120 is 100.0000 g. The initial internal weight value of
the GR-200, GR-300 and GR-202 is 200.0000 g.
200g 200g
Example
Weighing the same weight
The internal weight is corrected to
+0.6 mg. The balance is calibrated
with the internal weight.
Operation
1. Turn off the display using the [ON:OFF] key.
2. Press the [ON:OFF] key while the [RANGE] key and
[PRINT] key are pressed and held. Then the balance
displays PS.
3. Press the [PRINT] key. Then the balance displays the
switches.
4. Set the following switches to 1.
[RE-ZERO] key The key to select the setting of the switch.
[RANGE] key The key to select the switch to change the
setting.
Press and hold
these keys
Press the key
Press and
hold the key
To Next Page
xxx
Switch for the function table
Switch for the internal weight
5. Press the [PRINT] key to store the new setting.
The balance will return to the weighing mode.
6. Press and hold the [RANGE] key to enter the function table
and release the key when bASFnc is displayed.
7. Press the [RANGE] key several times to display CS in.