A&D GH Series Manuál s instrukcemi Strana 38

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 71
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 37
A&D GR Series Instruction Manual
GB 38
The [PRINT] key can be used at any time for transmitting data.
Key Mode
When you press the [PRINT] key and the display value is stable, the balance outputs the weighing data
and the display blinks one time.
Required setting dout Prt 0 Print key mode
Auto-Print Mode A
When the display value is stable and meets the conditions of “Auto-print polarity”, “Auto-print band” and
standard value (of zero point), the balance outputs the weighing data. If you press the [PRINT] key, the
balance outputs the data and the display blinks one time.
Required setting dout Prt 1 Auto-print mode A
dout AP-P Auto-print polarity
dout AP-b Auto-print band
Example “Weighing and removing one item.”
Auto-Print Mode B
When the display value is stable and meets the conditions of “Auto-print polarity”, “Auto-print band” and
standard value (of last stable value), the balance outputs the weighing data. If you press the [PRINT] key,
the balance outputs the data and the display blinks one time.
Required setting dout Prt 2 Auto-print mode B
dout AP-P Auto-print polarity
dout AP-b Auto-print band
Example “Transmitting the data of each operation.”
Stream Mode
The balance outputs the weighing data continuously.
Required setting dout Prt 3 Stream mode
dout dAtA 0 Data memory function is not used
bASF nc SPd Display update rate
SiF bPS Baud rate
Example “Monitoring data on a computer.”
Note: When the baud rate is set to 2400bps or less, the display update rate is faster than the
output rate and the balance may not transmit the data completely (and transmit intermittently).
9.1. “Data output mode”