8-2-2 Two-point Calibration
1 In the standby mode, press and hold
the HOLD key to enter the
calibration mode. " Cal " appears.
2 Select two-point calibration (Cal-2)
and press the PRINT key to confirm.
The standby mode of the two-
calibration mode appears.
Use the MODE key to switch
between the calibration modes,
one-point calibration (Cal-1) or
two-point calibration (Cal-2).
* Press the STOP key to exit the
calibration mode. The display
returns to the standby mode.
3 In two-point calibration mode, the calibration
mode ID indicator (-) appears below the
temperature display.
4 Place the standard viscosity fluid in the
sample cup. Press the START key to start
the measurement of the first point.
5 After the measurement, wait for the display
to become stable and press the PRINT key.
The measurement value blinks and is ready
to be corrected.
6 Correct the value using the following keys:
MODE key Switches the blinking digits.
START key Changes the setting of a
blinking digit.
STOP key Moves the decimal point.
7 Press the PRINT key to confirm the correction value.
8 To correct the value:
In the calibration standby mode to enter the
second point, press the STOP key and
correct the value.
Below is an example of the SV-10H.