5-4. Explanations about each function of the RsMulti
(1) The RsMulti window
Table
Records received data.
To change the title, double-click the title of the table (New.csv ).
Connection Name
List of setup connection names. Connects to a selected instrument.
Test / Monitor button
The connection can be checked in the “Test” display.
-Sending a send data command to the weighing instrument for data acquisition.
Press the [Test] button to send the command set in the <Property> window.
If a response to the command is displayed, communication is normally made.
-Sending data from the weighing instrument by pressing the [PRINT] key on the weighing
instrument.
(When used in the Key mode, Auto print mode and Stream mode.)
Blank out the command box and press the [Test] button in the <Property> window.
Press the [PRINT] key on the weighing instrument to send data.
If data is displayed, communication is normally made.
Start / Stop button
In the Start display, communication can be started by clicking this button.
In the Stop display, communication can be stopped by clicking this button.
Command button
To send a setup command, click this button.