AD-4408C Weighing Indicator Page 6
3.3. General
3.3.1. Interface
Standard serial output
Output for communication with A&D peripheral devices (20
mA current loop)
CC-Link interface CC-Link ver.1.10 remote device station
3.3.2. Measuring functions
Zero adjustment
• Sets the gross weight to zero by pressing the ZERO key or
a command from the CC-Link master station.
• Selection of disable or enable for the operation when
unstable.
• The zero value is battery backed up.
• Zero adjustable range
Can be set optionally in the range of 1 to 100% of the
weighing capacity.
Zero tracking
• Tracks the weight drift around the zero point to maintain
zero.
• Zero tracking time
0.0 to 5.0 seconds (Can be set optionally within the range.)
• Zero tracking band
0.0 to 9.9 d (Can be set optionally within the range.)
Tare
• Sets the net weight to zero by pressing the TARE key or a
command from the CC-Link master station.
• Selection of disable or enable for the operation when
unstable or negative.
• The tare weight is battery backed up.
• Tare range: Gross weight ≤ Weighing capacity
Stability detection
• Turns ON the stabilization indicator when the variables of
the weight values per sampling are within the set band in
the set time.
• Stability can be confirmed using the CC-Link.
• Stability detection time
0.0 to 9.9 seconds (Can be set optionally within the range.)
• Stability detection band
0 to 9 d (Can be set optionally within the range.)
Digital filter Cutoff frequency range: 0.07 to 11 Hz