GX-04K (Comparator output / RS-232C / Current loop output) GX-06K (Analog output / Current loop output)
8 Class Item and Parameter Description C5 in Internal mass correction Displayed only when the internal mass value correction switch is set to
9 4. Comparator Output (GX-04K) Comparator output, is the function to output the comparison results between the weighing data and upper/lower
10 4-2. Comparator (GX-04K) Specifications Maximum contact voltage: 100 VDC Maximum contact current: 100 mA DC Maximum contact resistance:
11 LO OK HILower limitUpper limitConducting(Short) Conducting (Short) Conducting(Short)Weighing data 4-3. Using The Comparator Output To use t
12 Class Item and Parameter Description Cp Hi Upper limit Cp lo Lower limit Refer to "4-4. Setting The Upper And Lower Limit Values" Dis
13 Entering the upper and lower limit values Step 8 With Cp Hi displayed, press the PRINT key. The current setting of the upper limit
14 Entering the reference and tolerance values Step 8 With Cp ref displayed, press the PRINT key. The current setting of the reference va
15 4-5. Example Of Use The following example uses the AD-8951 comparator light, which is sold separately, to display the comparison result in
16 5. Serial Output 5-1. RS-232C (GX-04K) Specifications D-Sub 25 pin numbers D-Sub 25 pin assignments Pin No. Signal name Interface type
17 RS-232C Terminals External contact input By connecting pin 18 (PRINT command) to pin 7, or pin 19 (RE-ZERO command) to pin 7
This Manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Ins
18 5-2. Current Loop Output (GX-04K/GX-06K) Specifications The specifications of the current loop interface are as follows: Transmission system
19 1 2 3 4MODEON1 2 3 4MODEON 5-3. Connection Of The AD-8121B Printer Preset the following parameters to use the AD-8121B printer. Class
20 5-4. Description Of The Item “Data output mode” The parameter setting of "Data output mode (prt)" applies to the performance when t
21 Interval memory mode The weighing data is periodically stored in memory. Example For periodical weighing without a personal computer co
22 KF format 5if type 2 This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only communicate using this
23 5-6. Description Of The Data Format Added To the Weighing Data ID number dout 5-id 1 The number to identify a specific balance. This form
24 5-7. Data Format Examples Stable Unstable Overload Positive error Overload Negative error Space, ASCII 20h Carriage
25 Units A&D D.P. KF MT g kg Counting mode Precent mode Ounce (Avoir) Pound Pound Ounce Troy Ounce Metric Carat Momme Pennyweight Grain Tael
26 5-8. Connection To A Computer And The Use Of WinCT The balance is of the DCE type (Data Communication Equipment), which can be connected to a
27 5-9. Commands Note A command has a terminator added, that is specified using "5if Crlf" of the function table, and is sent to the b
Instruction manual 1 GX-K series, GF-K series Contents 1. Introduction ...
28 Commands to control the balance UN:mm Recalls the unit mass values in memory. mm: 01 to 50. UW:∗∗∗∗∗∗.∗~~g Changes the unit mass value. Use "
29 5-11. Control Using CTS And RTS Depending on the "Ct5" parameter of "Serial interface (5if)", the balance performs as
30 6. Analog Output (GX-06K) 6-1. Installing The GX-06K Install the GX-06K option in the following procedures: 1. Remove two screws. Remove t
31 6-2. Analog Output (GX-06K) Specifications The specifications of the analog output unit (GX-06K) are as follows: Output impedance 100 Ω or
32 6-3. Function Table Of The Analog Output The “Analog output (aout)” of the function table can be selected when the GX-06K option is installed
33 Example When an 0 is set: Analog output voltage:"0V~" 0 V to 1 V 5el 0 0.67 V 5el 1 0.56 V 5el 2 0.45 V 5el 3 0.34 V
34 6-4. Switching Output Voltage The output voltage can be switched using the slide switch on the GX-06K panel. "0V~" has been set at f
35 6-6. Fixed Output Voltage The output voltage is fixed under the following conditions: 1. During operations other than weighing : 0 V (
2 1. Introduction This manual describes how the GX-K series and GF-K series options, GX-04K and GX-06K work, and how to get the most out of them
3 Comparator output Contact-outputs, output the comparison results between the weighing data and upper/lower limit values, using HI, OK, and L
4 3. Function Table The function table reads or rewrites the parameters that are stored in the balance. When GX-04K or GX-06K is used, set the fu
5 3-2. Details Of The Function Table Class Item and Parameter Description 0 Fast response, sensitive value FAST 1 MID.Cond Condition 2 Slo
6 Class Item and Parameter Description Cp Hi Upper limit Cp lo Lower limit Refer to "4-4. Setting The Upper And Lower Limit Values" Disp
7 Class Item and Parameter Description 0 No pause pU5e Data output pause 1 Pause (1.6 seconds) Selects the data output interval. 0 Not used